Answers are grounded in this paper and cite their sources.
Purpose: To determine the glycerin concentration that best maintains the volume of the 5-cc and 10-cc balloons of the Duette double-balloon chronic indwelling catheter (DBCIC). Materials and Methods: Over the course of 2 experimental trials, 16 Fr double-balloon catheters were filled with glycerin solutions of various concentrations (0%-40% v/v in sterile water) and normal saline and placed in synthetic urine. Balloon diameters were recorded over time, and final balloon volumes were measured on day 28. Results: The 30% and 20% glycerin solutions were most effective at minimizing volume change of the 5-cc and 10-cc balloons, respectively, in both trials. Furthermore, linear regression analysis of volume change (ΔV) vs glycerin concentration revealed minimization (ΔV = 0) for the 5-cc balloon occurred at 27% glycerin in the first trial and 30% glycerin in the second trial. Minimization for the 10-cc balloon occurred at 16% glycerin in both the first and second trials. Conclusions: For the sake of simplicity and ease of performing calculations and dilutions in a busy clinical environment, we recommend that 30% glycerin be used as a filling solution for the 5-cc balloon and 15% glycerin be used as a filling solution for the 10-cc balloon of the DBCIC in vivo. Further studies conducted in vivo are needed to investigate the effectiveness of these filling solutions in clinical practice.
No takes yet. Share an insight, caveat, or question.
Vinayak et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: