Answers are grounded in this paper and cite their sources.
Background: Robot-assisted spine surgery depends on rigid fixation of the robotic system reference frame to the posterior superior iliac spine or spinous processes of the patient. However, if the spinous processes are absent or are fragile, we propose and have successfully utilized a reference clamp anchoring technique affixed to a rod that transversely connects the two pedicle screws of the same vertebra. Methods: We have described and have utilized the clamp anchoring technique to perform robot spine surgery. Results: In a previously operated thoracic segment, the rod-based reference provided stable/successful anchoring sufficient to achieve accurate registration for robot-assisted pedicle screw placement without complications. Conclusion: In the absence or fragility of spinous processes, anchoring the reference frame to a rod bridging two pedicle screws of a single vertebra successfully achieved accurate intraoperative registration to proceed with robotic spine surgery.
No takes yet. Share an insight, caveat, or question.
Defino et al. (2026) studied this question.