Key result
Sim&Size virtual simulation is linked to ~13-point higher technical success for Artisse intrasaccular device sizing.
Why the study?
Virtual simulation is increasingly promoted for intrasaccular device sizing, but whether it changes device selection or procedural outcomes for new, highly conformable devices such as the Artisse 2.0 was unknown.
Does Sim&Size virtual simulation improve device selection and procedural technical success compared to IFU chart-based sizing for the Artisse 2.0 intrasaccular device?
Cohort (n=123)
Yes
Does Sim&Size virtual simulation improve device selection and procedural technical success compared to IFU chart-based sizing for the Artisse 2.0 intrasaccular device?
Answers are grounded in this paper and cite their sources.
Sim&Size virtual simulation and IFU chart sizing selected the same Artisse 2.0 device in over 90% of cases, though simulation use was associated with higher overall technical success, particularly in anatomies at the limits of the device range.
BACKGROUND AND PURPOSE: Virtual simulation is increasingly promoted for intrasaccular device sizing, but whether it changes device selection or procedural outcomes for new, highly conformable devices is unknown. We compared Sim&Size simulation with instructions-for-use (IFU) chart sizing for the Artisse 2.0 in early multicenter experience. MATERIALS AND METHODS: We retrospectively analyzed data from the INSPIRE-A registry (November 2022-June 2024). Sim&Size use depended on software availability. We assessed agreement between sizing tools, technical success, first-device success, adjunctive devices, procedure and fluoroscopy times, complications, and the cause of each technical failure. RESULTS: Sim&Size was used in 88/123 cases (71.5%). When both methods were used (n=59), the same device was selected in 54 (91.5%). Technical success was 98.9% with Sim&Size and 85.7% without (p=0.007), whereas first-device success (86.4% vs 80.0%, p=0.41) and procedure and fluoroscopy times did not differ. Perioperative complications occurred in 4/123 (3.3%). Of the six technical failures, five were related to device fit, mostly in anatomies at the limits of the device range, and one resulted from a perforation before deployment. A dome-to-neck ratio <1 was associated with failure (3/7 vs 3/116, p=0.002). CONCLUSIONS: For the Artisse 2.0, Sim&Size and chart-based sizing selected the same device in most cases. Sim&Size use was associated with higher technical success, whereas first-device success and procedure and fluoroscopy times were comparable. Both approaches achieved high procedural success with a low complication rate and may be used according to operator preference, with simulation potentially most useful in anatomies at the limits of the device range.
No takes yet. Share an insight, caveat, or question.
Rodriguez‐Erazú et al. (2026) conducted a cohort in Intracranial aneurysms treated with the Artisse 2.0 intrasaccular device (n=123). Sim&Size virtual simulation sizing vs. Instructions-for-use (IFU) chart-based sizing without Sim&Size was evaluated on Technical success (p=0.007). Sim&Size virtual simulation for Artisse 2.0 sizing was associated with higher technical success than IFU chart sizing alone.
Synapse has enriched 10 related papers. Consider them for comparative context: