Key result
Radiologist preoperative EVAR measurements exceed company planning values by ~1.3 mm.
Why the study?
Concordance among preoperative EVAR measurements made by a radiologist, a vascular surgeon, and the device manufacturer's planning team, along with the association between intraoperative contrast volume and postoperative renal function changes, required evaluation.
Cohort (n=40)
No
Preoperative EVAR sizing measurements from radiologists, vascular surgeons, and device manufacturers are not fully interchangeable, supporting structured multidisciplinary reconciliation alongside routine postoperative renal biochemical surveillance.
Answers are grounded in this paper and cite their sources.
Measurement differences between radiologists and companies may affect EVAR planning; leaves open standardization needs in larger studies.
Aim: To assess concordance among preoperative measurements made by a radiologist, a vascular surgeon, and the device manufacturer's planning team for endovascular aneurysm repair (EVAR), and to explore the association between intraoperative contrast volume and postoperative change in renal function. Material & Methods: In this prospective, consecutive, single-centre observational study, 40 patients undergoing EVAR were assessed. Three independent sets of preoperative measurements were compared using Bland-Altman analysis, paired t-tests, and absolute-agreement intraclass correlation coefficients. Paired changes in renal function and sac diameter were assessed using Wilcoxon signed-rank tests; exploratory associations between contrast volume and changes in renal function were assessed using Spearman correlation; and all-cause survival was estimated using Kaplan-Meier analysis. Results: Radiologist measurements exceeded company measurements by a mean of 1.29 mm (95% limits of agreement -1.84 to 4.43 mm; ICC 0.80; n=34; p<0.001) and were 2.24 mm higher than surgeon measurements (n=21; p<0.001). The mean surgeon-company difference was -0.50 mm (n=20; p=0.404). Median paired eGFR values decreased by 5.0 mL/min/1.73 m² (n=26; p=0.017), while creatinine and urea increased by 5.5 µmol/L (n=30; p=0.002) and 1.2 mmol/L (n=17; p=0.003), respectively. Contrast volume was not significantly associated with changes in renal function (eGFR n=24, creatinine n=27, urea n=14; all p≥0.634). Across all recorded imaging, 17/40 patients (42.5%) had any endoleak, comprising 16 type II and one type Ib endoleak; no type III endoleaks were recorded. Paired sac-diameter measurements decreased significantly (n=20; p=0.018). Provisional overall survival was 94.4% at six months and 88.1% at both one and two years. Conclusion: Measurements from the three planning sources were not fully interchangeable, supporting structured reconciliation of discrepancies during EVAR planning. Postoperative renal deterioration was observed, but the exploratory analysis did not identify a statistically significant association with contrast volume. Continued renal and imaging surveillance is warranted, and larger multicentre comparative studies are needed.
No takes yet. Share an insight, caveat, or question.
Kordzadeh et al. (2026) conducted a cohort in Abdominal aortic aneurysm undergoing endovascular aneurysm repair (EVAR) (n=40). Preoperative EVAR diameter measurements by an interventional radiologist vs. Preoperative EVAR diameter measurements by the device manufacturer's planning team and vascular surgeon was evaluated on Mean bias in preoperative aneurysm diameter measurements (radiologist minus company planning team) (MD 1.29 mm, 95% CI 0.74 to 1.85, p=<0.001). Radiologist preoperative EVAR measurements exceeded company measurements by 1.29 mm and surgeon measurements by 2.24 mm, while post-EVAR eGFR declined by 5.0 mL/min/1.73 m².