To overcome the limitation of long calibration time of traditional infrared radiation calibration methods, the paper presents the efficient calibration method for wide dynamic range infrared radiation measurement systems. First, by analyzing the energy transfer mechanism in systems incorporating variable optical attenuator and adjustable integration time settings, we develop the comprehensive system response model that accounts for the effects of both optical attenuator and integration time, incorporating nonlinear corrections in both dimensions. The model is designed to be universally applicable and can be seamlessly integrated with the inner-outer calibration concept, enabling full calibration method coverage—from collimator-blackbody and close-range extended source to inner-outer calibration—within a unified efficient framework. Consequently, the efficient calibration algorithm is derived based on the model, utilizing nonlinear optimization for parameter estimation. Experiments demonstrate that the proposed method boosts calibration efficiency multiple times relative to traditional approaches Furthermore, compared to Yang's fast calibration method, the approach does not require strict linearity in the attenuation and integration time dimensions, making it applicable to a broader range of infrared radiometric systems.