Endotracheal tube (ETT) malposition may occur in children with congenital heart disease because of shorter tracheal length and growth restriction. This prospective diagnostic accuracy study evaluated suprasternal ultrasonography for confirming ETT depth in 116 preschool children (0–6 years) undergoing elective cardiac catheterization under general anesthesia. Microcuff ETTs were inserted using the vocal cord depth mark. Ultrasonographic visualization of the cuff at the suprasternal notch (SSN), a prespecified surrogate for correct tip position, was compared with fluoroscopic tip position as the reference standard. Fluoroscopy showed correct tip position in 99 patients (85.3%). Using cuff visualization at the SSN as the criterion, sensitivity was 40.4% (95% CI, 30.7–50.7), specificity 70.6% (95% CI, 44.0–89.7), and overall agreement 44.8%, with negligible agreement beyond chance. In post hoc exploratory analyses, extending the threshold to 1.1–1.3 cm above the SSN increased sensitivity to above 80% but reduced specificity to 20%. The size-dependent distance between the cuff and the tube tip may explain part of this discordance. The SSN-based cuff criterion has limited accuracy for confirming ETT depth in this population and should not be used alone; revised ultrasonographic criteria require prospective validation. Clinical trial registration The study was registered in the Thai Clinical Trials Registry before data collection process (TCTR20230218001).