Computed tomography (CT)-based three-dimensional (3D) morphometry is widely used to describe the proximal femur, yet the measurement properties of individual parameters are rarely reported. We quantified the inter-rater reliability, agreement and measurement error of ten proximal femoral parameters measured independently by two blinded raters on 3D models of 20 adult dry cadaveric femora. Reliability was quantified with the intraclass correlation coefficient [ICC(2,1)], measurement error with the standard error of measurement (SEM) and minimal detectable change, and agreement and systematic bias with Bland–Altman analysis and paired t-tests. Six linear dimensions and the distal canal diameter (FCD2) showed excellent reliability (ICC 0.936–0.995; SEM 0.32–1.31 mm). The neck-shaft angle showed good reliability (ICC 0.778), and femoral anteversion (FAA; ICC 0.676) and the proximal canal diameter (FCD1; ICC 0.705) moderate reliability. Significant systematic differences between raters occurred for FAA (4.55°), FCD1 (1.13 mm) and FCD2 (0.49 mm), the latter despite excellent reliability. FAA had the largest measurement error (SEM 4.90°; limits of agreement − 9.0° to + 18.2°). These results describe agreement between two raters under one protocol; they do not establish anatomical accuracy, which requires comparison with a reference standard.