Answers are grounded in this paper and cite their sources.
This study was aimed to evaluate the validity and agreement of MySmartOH (a novel cloud-based analytical software), as compared to conventional audiometric test, for screening of occupational noise–related hearing disorders (ONRHDs) among Malaysian workers. Subsequently, the time taken for screening and reporting between MySmartOH and the conventional paper-based approach was compared. This cross-sectional study involved 16 occupational health doctors and 320 audiograms. For each hearing condition, sensitivity and specificity tests were used to determine the software’s validity while Cohen’s kappa coefficient was used to assess its agreement. The time taken to screen and report ONRHDs between MySmartOH and the conventional method was analysed. The sensitivity of MySmartOH ranged from 55.6 to 99.1% while specificity ranged between 70.0 and 99.2%. Cohen’s kappa values for MySmartOH ranged between 0.41 and 0.92. For the left ear, screening and reporting were faster with MySmartOH (0.93 and 0.48 min) than with the conventional method (1.25 and 0.87 min). For the right ear, MySmartOH also shortened screening (0.87 vs. 1.15 min) and reporting times (0.47 vs. 0.83 min). MySmartOH reduced screening time by 24.3–25.6% and reporting time by 43.3–44.8%. MySmartOH demonstrated variable diagnostic performance, with moderate-to-high agreement for screening ONRHDs compared with conventional audiometric assessment. It also offers a faster alternative to the conventional method.
No takes yet. Share an insight, caveat, or question.
Pathmanathan et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: