Mechanically scanned linear arrays provide a practical route to three-dimensional photoacoustic imaging, but a finite elevation response causes the same target region to contribute to several neighboring B-scans. A smaller scan step therefore increases sampling density without, by itself, removing elevation-direction mixing. Here, fine-step scans are formulated as overlapping observations governed by an experimentally measured post-beamforming elevation response. A 0.05 mm wire was used to identify a 15-point response for a 128-element, 20 MHz array scanned at 0.1 mm increments; this response was embedded in an extended-grid forward operator, and the resulting non-negative composite-regularized inverse problem was solved using the alternating direction method of multipliers. In an independent 0.08 mm crossed-wire experiment, the median elevation full widths at half maximum for delay-and-sum (DAS), Wiener, Richardson–Lucy, and ADMM reconstructions were 0.854, 0.665, 0.630, and 0.590 mm, respectively. ADMM achieved a median paired FWHM reduction of 26.3% relative to DAS, with a Peak SNR of 23.30 dB and a PSLR of −8.66 dB. Finger-vessel imaging showed reduced apparent broadening and structural merging. These results show that explicitly modeling experimentally identified inter-scan coupling can improve elevation localization using information already present in conventional fine-step scans without changing the transducer aperture.