Answers are grounded in this paper and cite their sources.
Hydrogen peroxide (H2O2) is both a biologically relevant redox species and a common transduction product in oxidase-catalyzed assays, making its concentration indicative of biological redox status or clinically important substrates such as glucose. However, most H2O2 detection methods produce continuously varying signals and therefore lack a distinct visual readout at physiologically relevant decision thresholds, limiting rapid point-of-care assessment. Here, Asc−Ti2C QDs with a ligand−framework architecture were designed to generate a visible readout at a defined H2O2 threshold. The contrasting reactivities of the ascorbate ligands and Ti2C framework produced an abrupt fluorescence transition, and this threshold could be systematically tuned by adjusting the ligand-to-framework ratio. As a proof-of-concept application, the Asc−Ti2C QDs were coupled with glucose oxidase, immobilized in a PVA/PVP matrix, and integrated into a cassette for hypoglycemia screening. The resulting device exhibited a readily visible fluorescence transition at 4.0 mM glucose, close to the 3.9 mM hypoglycemia alert threshold, enabling direct naked-eye identification without instrumental quantification. Overall, this work advances fluorescence sensing beyond continuous quantification toward direct readout of biologically relevant thresholds, enabling more rapid and actionable diagnostic decisions.
No takes yet. Share an insight, caveat, or question.
Hu et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: