Polymer monoliths have emerged as attractive stationary phases for peptide and intact-protein analysis, offering an alternative to packed-bed columns. Their synthesis involves several critical manual steps raising concerns about reproducibility and transferability of the synthesis protocol. This study systematically evaluates the robustness and reproducibility of an established synthesis protocol where two operators synthesized two of column batches of 100 µm i.d. poly(styrene-co-divinylbenzene) monolithic capillary columns and assessing their performance in peptide separations. Column robustness was investigated through 70 consecutive peptide separations performed on a representative polymer monolithic capillary column. The column maintained its chromatographic performance, achieving retention-time stability comparable to commercial packed nanoLC columns (CV < 0.5%). The within-batch reproducibility of the columns was found to be generally high, with the coefficient of variance (CV) of the pressure drop below 10% and CV of retention times and peak width below 3% and 8-15%, respectively. Multivariate testing by multivariate analysis of variance (MANOVA) was performed to compare batches of the same operator and columns from different operators, detecting no significant difference between the operators (p = 0.2802). Our study demonstrates the reproducibility of polymer monolith fabrication across multiple levels, establishing the robustness and transferability of the synthesis protocol.