This is the authors' abstract. We don't add key points for this paper.
This study describes the design and preliminary functional evaluation of an Internet of Things (IoT)-based system for monitoring maternal heart rate and peripheral oxygen saturation (SpO2). The prototype integrates a MAX30100 pulse oximeter, a NodeMCU ESP8266 microcontroller, an OLED display, and a Blynk-based interface for remote visualization. The system acquires pulse-related signals, processes the sensor output, and transmits the reported heart rate and SpO2 values through a Wi-Fi connection. The prototype was evaluated through component connection checks, functional tests, and data-transmission tests. The manuscript reports ten participant records and an approximately 5-s transmission delay; however, the available records do not include a documented comparison against a calibrated reference oximeter. Therefore, the participant data are treated as preliminary observations rather than evidence of clinical accuracy. In addition, the blood-pressure values previously presented in the manuscript are not retained as validated measurements because the documented hardware description does not explain a cuff-based measurement mechanism, pressure calibration, or an oscillometric algorithm. The study consequently focuses on IoT integration and preliminary monitoring functionality, while clinical accuracy and blood-pressure measurement require further validation.