Key result
Methylene blue administration stains ECMO return tubing, falsely mimicking membrane lung dysfunction despite normal oxygenation.
Why the study?
Methylene blue administration during V-A ECMO can cause dark discoloration of the ECMO return tubing that mimics membrane lung dysfunction.
Case Report (n=1)
No
Repeated methylene blue administration during V-A ECMO can stain flexible PVC return tubing and mimic membrane lung dysfunction, requiring post-membrane lung blood gas evaluation and inspection of unstained rigid connectors to avoid unnecessary circuit interventions.
Answers are grounded in this paper and cite their sources.
Methylene blue may mimic membrane lung failure on V-A ECMO; this case report leaves open whether routine post-membrane blood gas analysis prevents unnecessary interventions.
We report the case of a 72-year-old man who underwent cardiac surgery for Group B Streptococcus infective endocarditis with embolic hepatic abscesses. Severe postoperative vasoplegia was treated with repeated methylene blue boluses and infusions while on central veno-arterial extracorporeal membrane oxygenation (V-A ECMO) for refractory cardiogenic and vasoplegic shock. Following multiple methylene blue administrations, the ECMO return tubing appeared dark, raising concern for membrane lung dysfunction (MLD). However, post-membrane lung blood gases confirmed adequate oxygen exchange and visual inspection on the post-membrane lung side of the circuit revealed bright red blood within the membrane lung (ML) and all rigid connectors. The discoloration was determined to have resulted from methylene blue staining of the flexible PVC tubing which mimicked the appearance of the post-ML flexible PVC tubing in a failed membrane lung despite the oxygenator functioning appropriately. Awareness of this artifact is critical to avoid unnecessary interventions; evaluation of post-ML blood gases and visual comparison with unstained ECMO circuit components can distinguish this benign finding from true MLD.
No takes yet. Share an insight, caveat, or question.
Donaldson et al. (2026) conducted a case report in Severe postoperative vasoplegia and refractory cardiogenic shock on central V-A ECMO following cardiac surgery for Group B Streptococcus infective endocarditis (n=1). Methylene blue boluses and infusions during V-A ECMO was evaluated on Apparent membrane lung dysfunction (dark discoloration of ECMO return flexible PVC tubing). Repeated methylene blue administration in a 72-year-old man on V-A ECMO stained the flexible PVC return tubing, mimicking membrane lung dysfunction despite normal oxygen exchange.
Synapse has enriched 10 related papers. Consider them for comparative context: