Polymer ligating clips are widely used for vessel and tissue ligation in robotic surgery. This prospective, randomized, phase III noninferiority trial evaluated the efficacy and early safety of VasoClip, a polymer clip with a Z-groove and elastic notch, compared with WeckClip. Forty patients undergoing robotic urologic surgery were randomized 1:1 to VasoClip ( n = 20) or WeckClip ( n = 20). The primary endpoint was ligation success assessed per clip, with patient-level clustering accounted for using generalized estimating equations. A total of 573 clips were applied (300 VasoClip and 273 WeckClip). Overall success was 96.0% versus 92.4%, with a clustering-adjusted absolute risk difference of 3.6% (95% CI, -1.5% to 8.7%). The estimate met the prespecified − 10% noninferiority criterion; however, because the trial included only 40 patients and the original sample-size calculation did not account for within-patient clustering, this result should be interpreted cautiously. In the exploratory body-structure subgroup, success was numerically higher with VasoClip (94.9% vs. 86.1%), but the difference was not statistically significant ( P = 0.078). No Clavien-Dindo grade III or higher complications occurred, and no early device-related safety signals were observed during the two-week follow-up.