In this paper, a novel non-invasive blood glucose monitoring system based on microwave sensing using a compact, contactless and sensitive dual-coupled sensor configuration is proposed. The proposed sensor consists of two electromagnetically coupled resonators, configured as a bandpass filter centered around 2.2 GHz. The sensor is fabricated on a 40 × 40 mm 2 Rogers RO3003C substrate with a 20 mm gap between the resonators to accommodate different finger sizes and incorporates a dedicated mechanical isolation structure to minimize measurement variability caused by direct contact, applied pressure, and positioning uncertainty. The operation of the sensor is based on near-field electromagnetic coupling between the resonators and biological tissue, enabling detection of glucose-correlated variations in dielectric properties without direct skin contact. The experimental trials for glucose concentrations from 76 to 201 mg/dL showed a monotonic decrease in the transmission coefficient with increasing glucose concentration with a sensitivity of 0.031 dB/mg/dL and frequency shift sensitivity of 0.41 MHz/mg/dL. These results demonstrate the effectiveness of the proposed sensor and indicate the feasibility of sensitive contactless glucose monitoring using a microwave-based sensor.