Total ankle arthroplasty has advanced significantly over the past 2 decades, driven by improvements in implant design and surgical techniques. However, its success depends on accurate component alignment and positioning. This can be compromised by the parallax effect, a phenomenon in which an object appears displaced due to variations in viewing angles during fluoroscopy. This technical tip article presents intraoperative strategies to mitigate the parallax effect during total ankle replacement. Recommendations include proper limb positioning with the foot pointing upwards, ensuring the X-ray beam remains perpendicular to the surgical table, and using cushions to align the midcoronal plane. Techniques for both anteroposterior and lateral imaging are discussed, including the importance of knee elevation, centralization of the image field, and adduction of the limb. These measures aim to improve imaging accuracy and ensure optimal component placement. Recognition and prevention of parallax are essential to reduce the risk of implant malalignment and associated postoperative complications such as instability and early failure. By adhering to systematic fluoroscopy protocols, orthopedic surgeons can enhance intraoperative precision and improve patient outcomes. Level of Evidence: Level 5—expert opinion/technique tip.