To assess the analytical capabilities and educational potential of accessible instrumentation, we evaluated a 3D-printed, camera-based do-it-yourself (DIY) spectrophotometer. Wavelength mapping employed four reference light sources spanning 365-650 nm, and sodium copper chlorophyllin response was assessed at ten concentrations from 3 to 30 ppm, with three replicates per concentration. Spectra of five aqueous chromophoric samples were also compared with measurements from a commercial UV-Visible spectrophotometer. Mean digital intensities at 3 and 30 ppm were 159.99 ± 2.58 and 163.20 ± 2.04, respectively, and seven of ten concentrations exhibited coefficients of variation below 3%, yet no reproducible positive apparent absorbance maxima or systematic concentration trend emerged. Across the five samples, only 30.9-52.9% of assessed spectral positions satisfied exploratory quality-control criteria, and none within ±2 nm of the commercial absorption maxima passed screening, preventing reliable peak comparison. The prototype therefore provides a potential educational platform for examining spectral acquisition and instrumental limitations, but is not a validated quantitative spectrophotometric method.