Answers are grounded in this paper and cite their sources.
The triple point of carbon dioxide (CO 2 TP) is one promising candidate for the alternative of the mercury triple point (Hg TP). The long-term stability and consistency realizing the CO 2 TP has the essential importance on the feasibility of the alternative. The experimental study described here addresses this point. In order to exclude unessential influences, we selected a single thermometer of high stability to be used in this work, spanning four years. We have constructed three cells for this study. During a four-years span, the realization of the CO 2 TP using the first cell demonstrates a long-term stability of 0.04 mK and a standard deviation of 0.04 mK. The realizations by the other two cells differ from that of the first cell by 0.02 mK and 0.08 mK, respectively. The standard deviations of measurement are less than 0.065 mK at F=0.5, and 0.099 mK at F=1.0 for the other two cells. The regular weighing of the cells shows minor leaks from the sample in the cell. The cumulative leakage of the first cell in the four-years span is 52 g. Nevertheless, the long-term stability exhibits no systematic drift and independence of sample mass leakage. In summary, this work demonstrates a perspective of long-term stability and consistency with the realizations of the CO 2 TP.
No takes yet. Share an insight, caveat, or question.
Liang et al. (2026) studied this question.