Answers are grounded in this paper and cite their sources.
A toy Monte Carlo study of energy reconstruction for an inverted-coaxial point-contact germanium detector. All programs were written by a commercial large language model in a dialogue with the author, who proposed the methods and judged the results; no data of any experiment were used. The study compares the trapezoid, cusp and optimum linear filters with filters and charge-collection corrections found by symbolic regression with the single operator eml(x, y) = exp(x) − ln(y), with a polynomial and a neural network as controls, and reports learning curves and tests against artificial sharpening of peaks. Two further tests are included: eml expressions distilled from a neural network, and the energy obtained from a small causal transformer that predicts the next sample of the pulse. The note lists what the model implemented at once and where it was wrong. The complete code and the numerical results are attached. Text and figures: CC BY 4.0; code: MIT licence (see README).
No takes yet. Share an insight, caveat, or question.
M. Misiaszek (2026) studied this question.