Purpose To assess associations of preoperative scotopic pupil diameter with early postoperative vault reduction and 3-month vault after implantable collamer lens (ICL) implantation. Methods The study included 30 patients (60 eyes) undergoing EVO ICL implantation. Pupillary diameter and light responses were assessed preoperatively and at 1 week and 1 month. Central vault was measured by anterior segment optical coherence tomography at 1 day, 1 week, 1 month, and 3 months. Early vault reduction was defined as day-1 minus week-1 vault. Associations were examined using Pearson correlations and multivariable linear regression adjusted for anterior chamber depth and Oversizing, with HC3 robust inference. Results Mean vault decreased from 915.9 µm at day 1 to 695.8 µm at month 3. Preoperative scotopic pupil diameter was positively correlated with 3-month vault (r = 0.346; P=.007) and inversely correlated with first-week vault reduction (r = − 0.371; P=.003). After adjustment, each 1-mm larger scotopic pupil was associated with a 100.7-µm higher 3-month vault (HC3 95% CI, 25.4 to 175.9; P=.010). Model adjusted R² was 0.183. Mesopic pupil diameter decreased by 0.451 mm at week 1 (P<.001). Conclusions Preoperative scotopic pupil diameter was associated with higher 3-month vault and less early vault reduction. Pupillometry may complement anatomical assessment during ICL planning, although its value for clinical sizing requires further validation.