Answers are grounded in this paper and cite their sources.
Background Incorrect system configuration can compromise medical imaging metadata and disrupt access to diagnostic information. Objective To examine the workflow and patient-safety implications of a system time error in an ultrasound device. Methods This qualitative case report analysed one anonymised incident from the Swedish ReidarMTP repository using patient-safety and HIT classification, imaging workflow analysis, and a risk–control matrix. Results Incorrect system time resulted in inaccurate timestamps on echocardiographic images, hindering retrieval and disrupting workflow. Although no direct patient harm was reported, the retrieval difficulties created a potential risk of delayed access to diagnostic information. The original investigation considered BIOS battery degradation the most probable technical explanation but did not independently confirm it; unintended manual alteration of the system date/time remained an alternative hypothesis. The incident also highlighted gaps in routine time verification, administrative access controls, and automated alerts. Conclusion This case demonstrates how a technical timekeeping failure can compromise metadata integrity and create downstream patient-safety risks. The contribution lies in integrating incident classification, workflow mapping, and risk–control analysis to connect a technical failure with its operational consequences and potential safeguards.
No takes yet. Share an insight, caveat, or question.
Rehan et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: