Key result
Miniaturizing slipstream and tip detection to a 7-F transradial system using the SIDEPASS dual-lumen catheter and AnteOwl IVUS enabled successful IVUS-guided wiring of a left circumflex #15 CTO.
Why the study?
This report sought to demonstrate that combined slipstream and tip detection can be miniaturized to a 7-F radial system using the SIDEPASS short-monorail dual-lumen catheter and AnteOwl IVUS for a CTO lesion.
Can the combination of the slipstream and tip detection techniques be miniaturized to a 7-F transradial system using the SIDEPASS dual-lumen catheter and AnteOwl IVUS for CTO PCI?
Case Report
Can the combination of the slipstream and tip detection techniques be miniaturized to a 7-F transradial system using the SIDEPASS dual-lumen catheter and AnteOwl IVUS for CTO PCI?
Answers are grounded in this paper and cite their sources.
Miniaturizing the slipstream and tip detection techniques to a 7-F system allows a smaller radial access footprint during complex CTO PCI while maintaining advanced IVUS-guided wiring capabilities.
OBJECTIVE: To demonstrate that the combination of the slipstream and tip detection can be miniaturized to a 7-F radial system using the SIDEPASS short-monorail dual-lumen catheter and AnteOwl intravascular ultrasound (IVUS) for a chronic total occlusion (CTO) lesion. KEY STEPS: Treat the in-stent restenosis in left circumflex artery #14 and insert the AnteOwl IVUS over the same wire (first-wire lumen only) to delineate the spatial relationship between the stent struts and the jailed ostium of left circumflex artery #15 CTO. Advance the SIDEPASS short-monorail dual-lumen catheter over the same wire within a 7-F guiding catheter, pass a second guidewire through an optimal stent-strut cell into the CTO, and complete lesion preparation. POTENTIAL PITFALLS: Narrow clearance within a 7-F guide, first-wire instability when removing the dual-lumen catheter, and inaccurate ostium identification can hinder the procedure. TAKE-HOME MESSAGES: Miniaturizing slipstream and tip detection to a 7-F system provides a smaller radial access footprint for complex CTO percutaneous coronary intervention while preserving advanced IVUS-guided wiring.
No takes yet. Share an insight, caveat, or question.
Yoshinaga et al. (2026) conducted a case report in Chronic total occlusion (CTO) in left circumflex artery #15 with in-stent restenosis in left circumflex artery #14. 7-F transradial slipstream and tip detection using SIDEPASS short-monorail dual-lumen catheter and AnteOwl intravascular ultrasound (IVUS) was evaluated on Successful IVUS-guided wiring and lesion preparation of the chronic total occlusion. Miniaturizing slipstream and tip detection to a 7-F transradial system using the SIDEPASS dual-lumen catheter and AnteOwl IVUS enabled successful IVUS-guided wiring of a left circumflex #15 CTO.