Key result
Target Tetra detachable coils achieve ~73% immediate complete occlusion in small intracranial aneurysms.
Why the study?
Coil embolization of small (<= 5 mm) intracranial aneurysms remains technically challenging. The Target Tetra Detachable Coil features a tetrahedral configuration designed to facilitate stable coil framing and improve packing stability, prompting prospective evaluation of its safety and efficacy.
Does endovascular coiling with Target Tetra Detachable Coils improve immediate angiographic occlusion rate and packing density in patients with small (≤ 5 mm) intracranial aneurysms?
Observational (n=95)
No
Does endovascular coiling with Target Tetra Detachable Coils improve immediate angiographic occlusion rate and packing density in patients with small (≤ 5 mm) intracranial aneurysms?
Endovascular coiling with Target Tetra Detachable Coils achieved high rates of immediate and 12-month complete angiographic occlusion (72.7% and 89.7%, respectively) with low complication rates in small (≤ 5 mm) intracranial aneurysms.
Answers are grounded in this paper and cite their sources.
May inform coil selection for small aneurysms; leaves open the need for randomized trials to confirm durability.
Coil embolization of small (≤ 5 mm) intracranial aneurysms (IAs) remains technically challenging. The Target Tetra ® Detachable Coil (TTDC) features a tetrahedral configuration designed to facilitate stable coil framing and improve packing stability. Therefore, we prospectively evaluated the safety and efficacy of TTDCs for small IAs. In this prospective, single-arm, single-center registry study, patients with small IAs were enrolled between January 2024 and March 2025. Endovascular coiling was performed using TTDCs comprising ≥ 70% of the total coil volume. Primary outcomes were the immediate angiographic occlusion rate and packing density. Secondary outcomes included device-related events, periprocedural complications, and clinical and radiological outcomes at 6 and 12 months. Ninety-five patients with 99 IAs were enrolled, including 98 unruptured IAs and 1 ruptured IA (72.6% female; mean age, 59.5 ± 10.8 years). The technical success rate was 100%. The mean packing density was 32.2 ± 8.7%, with TTDCs comprising 89.5 ± 11.5% of the total coil volume. Wide-neck aneurysms accounted for 98.0% of cases; stent-assisted coiling was used in 34.3%. Device-related events occurred in 7 coils (2.6% of 271 used coils), including failure to load into the microcatheter ( n = 5) and difficult retrieval ( n = 2). Immediate complete occlusion was achieved in 72.7% of cases and adequate occlusion in 92.9%. At 6 and 12 months, complete occlusion rates were 93.8% and 89.7%, respectively, with no retreatment required. Complications included 1 intraprocedural rupture without sequelae and 2 symptomatic thromboembolic events, one with persistent morbidity at 12 months. There was no procedure-related mortality. The favorable occlusion rates and acceptable complication profiles suggest that TTDCs are safe and effective for treating small IAs.
No takes yet. Share an insight, caveat, or question.
Chung et al. (2026) conducted an observational in Small (≤ 5 mm) intracranial aneurysms (n=95). Endovascular coiling with Target Tetra Detachable Coils (TTDC) was evaluated on Immediate angiographic complete occlusion rate and mean packing density. Endovascular coiling with Target Tetra Detachable Coils for small intracranial aneurysms achieved 72.7% immediate complete occlusion and 32.2% mean packing density.
Synapse has enriched 10 related papers. Consider them for comparative context: