A fiber-optic photoacoustic gas sensor (FOPAS) that incorporates multi-frequency cooperative noise suppression and cross-orthogonal multiple-reflection enhanced excitation is proposed for simultaneous detection of trace dual-component gases. The optimized composite buffer chambers effectively suppress the acoustic noise at two independent resonance frequencies corresponding to the resonance tube and the cantilever. Compared with conventional cylindrical buffer chambers, the designed composite structure reduces the background signal by 23% and the noise level by 20%. The cross-orthogonal multi-reflection optical structure enables two excitation lasers to be individually coupled into separate fiber collimators and simultaneously injected into the sensor. This overcomes the inherent channel spacing limitations of conventional wavelength division multiplexers (WDMs) while avoiding wavelength-selective optical losses. The resonance tube and the cantilever exhibit independent resonance peaks at 938 Hz and 1216 Hz, respectively. Dual-gas signals are naturally decoupled by employing two independent resonance acoustic fields, whereby acoustic crosstalk between the channels and cross-interference between the target gases are rendered negligible. The minimum detection limits (MDLs) are 8.2 ppb for CO and 3.1 ppb for H 2 S. The corresponding normalized noise equivalent absorption (NNEA) coefficients are 2.04×10 −10 W cm −1 Hz −1/2 for CO and 2.49×10 −10 W cm −1 Hz −1/2 for H 2 S.