Answers are grounded in this paper and cite their sources.
Humphrey visual field (HVF)-equivalent virtual reality fields (VRFs) have shown promise in adult glaucoma. Research comparing HVF-equivalent VRF to standard automated perimetry (SAP) in pediatric glaucoma is limited. We aimed to evaluate association and agreement between HVF-equivalent VRF and SAP for children with glaucoma. This prospective study enrolled children with known or suspected glaucoma. Participants completed HVF, then VisuALL HVF-equivalent VRF at one visit. A masked ophthalmologist graded visual fields for defects and location. Global parameters were compared between devices. A point-by-point deviation plot analysis was performed. A total of 60 patients (116 eyes) were included, mean age 13.1 ± 2.93 years. Mean retinal nerve fiber layer thickness was 86.1 ± 21.20 μm. Average mean deviation (MD), pattern standard deviation (PSD), and foveal sensitivity significantly differed for HVF versus VRF. There was moderate association between HVF and VRF for MD ( P < 0.001) and PSD ( P < 0.001). Concordance by region for any defect was 76.2%; point-by-point match rate for total deviation plots was 75.8%. Test duration was longer for HVF versus VRF (4.17 ± 1.19 vs. 2.10 ± 0.18 min, respectively, P < 0.001). Most children (43/44, 97.7%) preferred VRF over HVF. HVF and HVF-equivalent VRF testing modalities yielded moderate association. HVF-equivalent VRF offers a shorter test that children preferred and may supplement SAP.
No takes yet. Share an insight, caveat, or question.
Tillit et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: