Background/Objectives: In exoscopic spine surgery, the exoscopic image, navigation, and reference imaging occupy separate monitors. We describe a surgeon-centered extended reality (XR) workspace that places these sources in a single consumer head-mounted display, and report its measured video delay and its use in one clinical case. Methods: The stereoscopic output of a 4K three-dimensional (3D) exoscope and the intraoperative computed tomography (CT) navigation display were streamed over a dedicated local network to an Apple Vision Pro, together with four locally rendered reference layers; the exoscopic image was delivered at 1920 × 1080 in side-by-side stereoscopic format, that is, 960 × 1080 per eye. The delay of each streamed image relative to its native display was measured at the mirrored output with a 240-frames/s high-speed camera. The system was used during exoscopic minimally invasive open-door laminoplasty through a 3-cm incision for multilevel cervical ossification of the posterior longitudinal ligament. Results: The head-mounted exoscopic image lagged the native monitor by 30.8 ± 2.3 ms (range, 29.2–33.3 ms; n = 5), and the navigation image by 49.2 ± 1.9 ms (range, 45.8–50.0 ms; n = 5). The headset was worn for approximately 100 min without adjustment, malfunction, or reversion to the fallback monitor, and the procedure was completed without complication. Conclusions: The configuration was technically feasible with a small, measured delay. No ergonomic, workload, or clinical benefit is claimed.