Slight myopia is often chosen during intraocular lens power selection to avoid a hyperopic result, but the appropriate offset depends on prediction uncertainty and the consequences of residual refraction. This analysis derived the target that minimizes expected loss under unequal penalties for myopic and hyperopic outcomes and assessed discrete lens choices. Under Gaussian prediction error and piecewise-linear loss, a -0.25 D target at a prediction standard deviation of 0.40 D implied a hyperopic-to-myopic loss-weight ratio of 2.8. A sensitivity analysis motivated by distance-acuity asymmetry assigned greater penalties to myopia, yielding hypothetical hyperopic targets of +0.17 to +0.39 D. With hyperopic outcomes weighted twice as heavily and refractive steps no larger than prediction standard deviation, choosing the available refraction nearest the continuous optimum added little expected loss. The analysis explains the tradeoffs implicit in habitual targeting; clinical application requires calibrated prediction errors and patient-appropriate loss functions.