Answers are grounded in this paper and cite their sources.
BackgroundLinear accelerators are widely used in radiotherapy. However, photoneutron production may occur during high energy photon radiotherapy. Neutrons have a high radiation weighting factor.PurposeTo compare the thermal neutron sensitive TLD responses with and without wedge use during simulated rectal radiotherapy using TLD600/700 pairs.MethodsTLD pairs were placed in the target (rectal region), bladder, and right femur regions of an anthropomorphic Rando phantom for simulated rectal radiotherapy. Computed tomography images were acquired and transferred to the Prowess treatment planning system (TPS). Wedge and non-wedge plans were created using 15 MV photons at a treatment dose of 2 Gy. Plans were delivered using a Siemens Artiste™ linear accelerator. TLDs were read using a Harshaw system, and Am-Be calibrated neutron sensitive responses were expressed in mSv/Gy. Wedge/non-wedge response ratios were also calculated.ResultsAm-Be calibrated values for wedge versus non-wedge plans were 0.37 vs. 0.19 mSv/Gy for the target, 0.43 vs. 0.17 mSv/Gy for the bladder, and 0.51 vs. 0.22 mSv/Gy for the right femur, corresponding to wedge/non-wedge ratios of 1.95, 2.53, and 2.32, respectively.ConclusionsNeutron sensitive TLD responses were higher for the wedge plan than for the non-wedge plan at the same measurement positions.
No takes yet. Share an insight, caveat, or question.
Çakır et al. (2026) studied this question.