Answers are grounded in this paper and cite their sources.
In this study, a green and efficient analytical method was developed based on modified gel electromembrane extraction (G-EME) combined with differential pulse voltammetry (DPV) using a modified carbon screen-printed electrode (CSPE) for the determination of diclofenac as the model analyte. The MOF-199@Cu-Co-LDH nanocomposite was synthesized and employed as a dual-functional modifier for both the agarose gel membrane and CSPE. Its hierarchical structure and accessible active sites enhanced mass transfer during extraction and improved the electrochemical response. Under optimized condition, the method exhibited wide linear dynamic ranges of 0.20–200.00 µg L −1 , 0.50–100.00, and 1.00–100.00 µg L −1 for diclofenac determination in water, urine and plasma samples, respectively. The limits of detection and quantification were in the ranges of 0.06–0.31 and 0.19–1.00 µg L −1 , respectively. Relative standard deviations (RSDs) ranged from 2.9% to 5.3%. The method was successfully applied to human urine and plasma samples, providing recoveries above 90% with RSDs below 5.2%. Integrating of G-EME with electrochemical detection eliminated the need for a separate desorption step and reduced the total analysis time to less than 20 min. Furthermore, the greenness of the method was evaluated using the Analytical GREEnness (AGREE) metric, yielding a score of 0.81, which confirms its environmental compatibility.
No takes yet. Share an insight, caveat, or question.
Niazipour et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: