Whether near-infrared (NIR) prediction intervals survive a change in instrument is rarely tested. Split-conformal coverage of five point estimators was measured on four paired transfer axes (corn, tablet, cross-manufacturer wheat, handheld soil units). Against a nominal 0.90 and 0.870 to 0.954 in domain, source-calibrated partial least squares (PLS) coverage fell to 0.022 on corn protein and 0.030 in the worse tablet direction; the most accurate models lost the most, whereas the handheld units lost little (PLS median 0.862). Covariate-shift weighting (corn only), difficulty-normalised scores and the leverage-only classical interval did not repair the collapse. Calibrating on labelled target samples (anchoring) restored validity. At the nine-sample floor, the coverage distribution followed the closed-form exchangeable law on corn and tablet, leaving 32% and 37% of calibrations below nominal (law: 31% and 38%). In a PLS tablet release decision, transfer passed 11 to 19 non-conforming tablets per hundred released (0.3 to 0.6 in domain; 21 at random). With fifteen samples, piecewise direct standardisation improved one handheld pair’s accuracy and harmed the other’s, whereas anchoring alone restored nominal median coverage on every axis. Nine to fifteen reference values measured on the new instrument restore a valid interval, to be reported with its spread.