Key result
High-frequency bipolar pulsed field ablation kills myocardial cells at ~720 V/cm while blood flow limits heating.
Why the study?
Because of the high voltage-to-electrode distance ratio in pulsed field ablation, thermal effects need to be considered when using numerical simulations to predict the ablation region for cardiac treatment planning.
Does accounting for dynamic blood flow and conductivity variations affect thermal distribution and ablation region predictions during high-frequency bipolar pulsed field ablation in H9C2 myocardial cells and pulmonary vein models?
Does accounting for dynamic blood flow and conductivity variations affect thermal distribution and ablation region predictions during high-frequency bipolar pulsed field ablation in H9C2 myocardial cells and pulmonary vein models?
Dynamic blood flow and conductivity effects significantly limit thermal rise during high-frequency bipolar pulsed field ablation and should be incorporated into computational treatment planning to assist in precise pulse selection.
Answers are grounded in this paper and cite their sources.
Blood flow must be modeled in PFA simulations to avoid thermal overestimation; extends in vitro thresholds to vein models but leaves open clinical translation.
Objective: Pulsed Field Ablation (PFA) is a promising non-thermal energy modality for catheter-based cardiac ablation. Treatment planning using numerical simulations of PFA is important for predicting the ablation region. However, owing to the high voltage-to-electrode distance ratio, thermal effects should be considered during treatment planning. This study aims to analyse the ablation region for cardiac treatment planning considering the thermal effect. Approach: this study aimed first to determine the lethal electric field threshold (LEFT) of H9C2 myocardial cells exposed to high-frequency bipolar pulses (HFBPs). Then two simplified pulmonary vein (PV) models with a circular electrode array were constructed to investigate electric field and thermal distributions, accounting for dynamic blood flow and conductivity variation. Main results: cell experimental results showed that under the condition of equivalent energized time, fewer pulses per burst and prolonged interpulse delay of HFBPs can reduce the LEFT of myocardial cells. The LET reached 720 V/cm when applying 120 HFBPs of 1000 μs duration. Numerical simulations revealed that temperature rises sharply when blood flow is neglected, with a maximum temperature of 67 ℃. However, thermal effects are substantially limited (peak temperature of 38.3 ℃) due to blood flow, even when HFBPs with a high voltage-to-electrode distance (1250 V) are applied. The temperature rise rapidly declines and reaches a plateau at higher blood velocities. HFBPs can produce a full circumferential ablation region, and the ablation width increases linearly with rising applied voltage. Significance: this study demonstrates that dynamic blood flow and conductivity effects should be considered for myocardial ablation assessments, which may assist clinicians in precise pulse selection.
No takes yet. Share an insight, caveat, or question.
Lv et al. (2026) studied Cardiac ablation. Pulsed Field Ablation using high-frequency bipolar pulses (HFBPs) vs. Neglected blood flow vs. dynamic blood flow in numerical simulation was evaluated on Lethal electric field threshold (LEFT) and peak temperature distribution. High-frequency bipolar pulses achieved a 720 V/cm lethal threshold in myocardial cells, while dynamic blood flow limited peak simulation temperatures to 38.3 °C vs 67 °C without blood flow.
Synapse has enriched 10 related papers. Consider them for comparative context: