Key result
Transfer function-derived central pressure yields similar baroreflex gains to invasive aortic catheterization.
Why the study?
Peripheral blood pressure underestimates pressure changes during baroreflex testing and overestimates baroreflex gain, while invasive central measurement presents ethical and practical problems. Generalized transfer function-derived central blood pressure from the peripheral pulse was evaluated as a potential alternative.
Does generalized transfer function-derived central blood pressure accurately estimate baroreflex gain compared to direct invasive aortic catheter measurement in patients undergoing cardiac catheterization?
Observational (n=22)
Does generalized transfer function-derived central blood pressure accurately estimate baroreflex gain compared to direct invasive aortic catheter measurement in patients undergoing cardiac catheterization?
Generalized transfer function-derived central blood pressure accurately calculates baroreflex gain only when calibrated invasively, whereas noninvasive calibration overestimates baroreflex gain.
Answers are grounded in this paper and cite their sources.
Invasive calibration may permit reliable baroreflex gain estimation via generalized transfer function; leaves open noninvasive calibration validity.
BACKGROUND: Peripheral blood pressure measurement underestimates pressure changes during baroreflex testing, resulting in an overestimation of baroreflex gain. This error might be reduced by measuring central blood pressure; the invasive measurement, however, may represent ethical and practical problems. The solution may be the derivation of central blood pressure from the peripheral pulse using a generalized transfer function. METHODS: In the current study, we tested the agreement between catheter-measured and generalized transfer function derived central blood pressure measurements and corresponding baroreflex gains. ECG and blood pressure waveforms were monitored continuously during a phenylephrine-induced pressure rise in 22 subjects undergoing cardiac catheterization. Pressure was measured with a catheter positioned in the aorta and with applanation tonometry in the radial artery. Radial pressure waveforms were subject to a generalized transfer function built in the SphygmoCor device to derive central pressure waveforms. Radial tonometric signal was calibrated with catheter-measured (invasive) and sphygmomanometric (noninvasive) pressures. Baroreflex gains were calculated from the linear regressions between heart period and systolic pressure changes. RESULTS: When radial tonometric signal was calibrated invasively, there was no group difference between baroreflex gains calculated from SphygmoCor-derived and catheter-measured pressures (8.2 +/- 1.2 vs. 7.2 +/- 1.2 ms/mmHg, P = NS). When radial tonometric signal was calibrated noninvasively, however, baroreflex gains calculated from SphygmoCor-derived pressures overestimated those calculated from catheter-measured pressures. CONCLUSION: Using a generalized transfer function is an accurate method to derive central pressure changes for baroreflex gain calculation. The technique, however, requires invasive pressure measurements for calibration, leaving the problem of a fully noninvasive central pressure measurement unresolved.
No takes yet. Share an insight, caveat, or question.
Studinger et al. (2008) conducted an observational in Subjects undergoing cardiac catheterization (n=22). Generalized transfer function-derived central blood pressure via radial artery applanation tonometry (SphygmoCor) vs. Invasive aortic catheter-measured central blood pressure was evaluated on Baroreflex gain calculated from linear regressions between heart period and systolic pressure changes (ms/mmHg) (p=NS). Generalized transfer function-derived central pressure yielded baroreflex gains similar to invasive aortic catheter measurements when calibrated invasively.