Implant-associated infections remain a major clinical challenge, motivating the development of antimicrobial surface coatings that can prevent bacterial colonization while supporting tissue integration. Here, titanium surfaces were functionalized with polydopamine (PDA) and subsequently coated with three synthetic antimicrobial peptides (GN-2, GN-4, and IDR-1018) or chlorhexidine (CHLX). The antimicrobial activity of the resulting surfaces was evaluated against pathogens commonly associated with implant infections, and their biocompatibility was assessed using human osteoblasts. Although all three peptides displayed strong antimicrobial activity in solution, their efficacy was reduced following surface immobilization. In contrast, CHLX retained substantial antimicrobial activity when immobilized via PDA and outperformed the peptide-coated surfaces. PDA-functionalized surfaces and peptide coatings showed good compatibility with human osteoblasts. These findings highlight the importance of considering how surface immobilization influences antimicrobial function and provide insights for the design of infection-resistant biomaterial surfaces for medical implants.