Answers are grounded in this paper and cite their sources.
This study aimed to determine the reliability and validity of the Performanz Spark for measuring sprint performance. Seventy-nine male basketball players completed two maximal sprints over two separate days, assessed using both the Performanz Spark and Witty photocell gates. The Performanz Spark demonstrated excellent intra- and inter-session reliability across all split times (ICC: .96-.98; CV: 1.23–2.02%; r: .95-.98). Absolute reliability was confirmed with low SEM (0.02–0.08 s; 1.18–2.87%) and MDC (0.06–0.22 s; 3.26–7.44%) values. Furthermore, excellent agreement was found between the Performanz Spark and Witty (ICC: .97-.99; CV: 1.23–1.95%; r: .94-.99), and Lin’s CCC values ranged from .94 to .98 (95% CI: .91-.99), indicating high accuracy and precision. Bland-Altman analysis revealed small but statistically significant positive systematic biases (0.006–0.007 s, p < .01) across all split times, indicating a slight but consistent overestimation of sprint times compared to the Witty. However, the very low R2 values (R2 ≤ .01) suggest no proportional bias, meaning these minimal differences were independent of sprint velocity. Despite this, the narrow limits of agreement and minimal mean differences confirm that the Performanz Spark is a valid and reliable tool for assessing linear sprint performance.
No takes yet. Share an insight, caveat, or question.
Seçer et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: