Objectives: The objectives of the study are to analyze internal quality-control results for mean corpuscular volume, mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC) using Westgard rules and Six Sigma and to compare analytical performance between two type C hospital laboratories. Materials and Methods: This retrospective quantitative descriptive study used secondary daily quality-control data from Hospital X and Hospital Y from November 2025 to March 2026. All eligible low-, normal-, and high-level control results were included. Statistical analysis: The analysis included mean, standard deviation, bias, recovery percentage, coefficient of variation (CV), Levey-Jennings charts, Westgard rule events, and Six Sigma values. Results: Most bias and recovery values met the pre-specified acceptance limits. The largest deviation occurred in the high-level MCHC control at Hospital Y, with a bias of 19.18% and recovery of 119.18%. Fifty-three Westgard rule events were identified: 52 1-2S warnings and one 2-2S rejection. Mean sigma was 6.47 at Hospital X and 4.96 at Hospital Y. MCH showed the best performance at Hospital X (sigma 8.53), whereas MCHC showed the lowest performance at Hospital Y (sigma 3.40). Conclusions: Analytical performance differed by erythrocyte index and laboratory. Hospital X showed more stable analytical quality, particularly for MCH, whereas MCHC at Hospital Y required closer quality-control monitoring.