Quantitative scanning thermal microscopy (SThM) uses different temperature references for thermocouple and thermoresistive measurements, limiting comparison and cross-validation. Here, we establish the first junction-referenced dual-mode SThM methodology that calibrates and characterizes both responses of a tungsten-chromium (W-Cr) ultra-thin oxide tip (UTOT) probe against a common junction-temperature-rise reference. Vacuum null-point SThM calibrates the room temperature Seebeck coefficient, which, together with the initial voltage-power slope, determines the low-power effective thermal resistance. The high-temperature axis uses the first-order approximation T j ≈ R p P . We determine effective thermal resistance, Seebeck coefficient, thermal time constant, electrical resistance, and junction-referenced effective temperature coefficient of resistance (TCR) in air and vacuum. Practical power-sweep limits correspond to estimated junction-temperature rises of approximately 300°C in air and 400°C in vacuum. R ( T j ) maps the thermoresistive response onto the common junction-temperature axis, enabling comparison of both readouts and providing a basis for cross-validation.