Answers are grounded in this paper and cite their sources.
Background: Point of care ultrasound (POCUS) measurement of the optic nerve sheath diameter (ONSD) has been established as a non-invasive method for estimating intracranial pressure in critically ill patients. However, the accuracy of the technique depends strongly on proper image acquisition. Errors in system optimization may result in inaccurate measurements and increased inter-operator variability. Objective: Our objective was to describe the most common technical errors in ONSD measurement and to propose a checklist to optimize image acquisition. Methods: A narrative review of the literature and an analysis of technical errors observed during POCUS training in critical care settings were completed. Emphasis was on ultrasound system parameters and image acquisition. Results: Six key technical elements were identified to optimize ONSD measurement. They included preset selection, depth adjustment, focal zone positioning, transducer frequency, gain optimization, and identification of the central retinal artery and vein. Conclusion: Standardization of ultrasound system settings may improve image quality and reduce measurement errors in ONSD, increasing reproducibility in emergency and critical care settings.
No takes yet. Share an insight, caveat, or question.
Bañuelos-Huerta et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: