Accurate extrinsic calibration is essential for combining 3D LiDAR measurements with camera images, especially when sparse laser sampling complicates feature extraction. This study proposes a low-cost target-based method for 3D LiDAR and monocular or stereo camera calibration. A passive planar target with six circular holes provides corresponding LiDAR and image features. Both variants share the acquisition arrangement and LiDAR feature extraction procedure. Initial transformations are estimated from 3D-2D correspondences for the monocular variant and triangulated 3D-3D correspondences for the stereo variant, followed by refinement minimizing image reprojection errors. Experiments with a Velodyne VLP-16 and ZED2 stereo camera included ten independent calibration datasets and analyses of target distribution. Mean rotational deviations across repeated calibrations were 0.059° and 0.052° for the monocular and stereo variants, respectively. Their transformations differed by 0.030° in rotation and 2.0 mm in translation on average. Single-position calibration produced large errors at excluded target positions, whereas configurations spanning multiple distances yielded lower validation errors. Six positions covering the nearest and farthest distances achieved a mean validation RMSE of 3.58 pixels at the excluded intermediate distance. These findings demonstrate close correspondence between calibration variants and highlight the importance of spatial target distribution.