OBJECTIVES: To develop a slice-specific dose library and supporting interpolation scheme for proper estimation of organ doses from CT scans based on the UF/MSK mesh-type adult and pediatric phantom series. METHODS: Monte Carlo (MC) simulations were performed using the Particle and Heavy Ion Transport code System (PHITS) for selected adult female and pediatric male phantoms across a user defined CT source term. Dose profiles were calculated on a 1-cm slice basis and categorized by tissue type (distributed vs. non-distributed). Linear interpolation of slice doses within a height series was evaluated using multiple phantom subsets and compared against MC results. RESULTS: Slice-by-slice linear interpolation produced total organ dose estimates within 5% of MC values for most studied tissues, with all errors below 7%. Mean errors for abdomen, chest-abdomen-pelvis and cardiac scan ranges were also below 5%, and mean errors for individual slices were below 3%. The method performed well across various interpolation schemes, especially when phantoms were selected based on similar construction methods. CONCLUSIONS: This approach enables accurate, scalable organ dose estimation with reduced computational demands. The resulting dose library and interpolation method will be integrated into the MIRDct platform, a dosimetry tool for calculating total organ dose values across a wide range of CT parameters. MIRDct is freely available to the medical imaging community at MIRDsoft.org.