Answers are grounded in this paper and cite their sources.
Matrix‐assisted ionization (MAI) generates gas‐phase analyte ions through sublimation of a solid matrix without applying voltage or laser energy. Despite growing applications, the proton source responsible for generating protonated analyte ions has not been unambiguously identified. In this study, hydrogen–deuterium exchange (HDX) experiments were performed using a home‐built MAI source coupled with a quadrupole mass spectrometer. By systematically replacing the matrix, solvent, and vapor molecules with their deuterated counterparts, we demonstrate that the protons on the detected analyte ions are exchanged with gas‐phase water molecules in the intermediate pressure region. This indicates that the detected ions are products of secondary HDX reactions between preformed protonated ions and water vapor. The ion abundance was unchanged from dry N 2 to 60% relative humidity, showing that water vapor exchanges the protons of preformed ions without creating them. The replacement ratios for L‐arginine (97.2%) and bradykinin (82.3%) are remarkably high, establishing that MAI‐generated ions are fully desolvated, in sharp contrast to the solvated ions produced by electrospray ionization. These findings further reveal that rapid secondary reactions in the subatmospheric pressure region can overwrite the isotopic signatures of primary ionization products, complicating direct identification of the original proton source in MAI.
No takes yet. Share an insight, caveat, or question.
Chao et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: