Answers are grounded in this paper and cite their sources.
The baseline design of the thermal filters of the X-Ray Integral Field Unit instrument on board NewAthena consists of ultra-thin films of polyimide and aluminum. These filters shield the detector from out-of-band radiation while maintaining high detector efficiency in the soft X-ray regime. In this paper, we present the measured transmission curves of different polyimide/aluminum filters with differing polyimide thicknesses across the X-ray to far-infrared spectral regions. We also compare the derived X-ray transmission of different combinations of polyimide thicknesses that may compose the stack. Our data show that: (i) the out-of-band rejection of a single filter is independent of polyimide thickness; (ii) the out-of-band rejection, nominally achieved with 30 nm of aluminum per filter, shows some variability that is likely due to aluminum coating defects; and (iii) under the hypothesis of a nominal 30 nm Al thickness per filter, a total polyimide thickness below 300 nm for the full stack complies with the in-band response requirements. Finally, the transmission spectroscopy test campaigns performed and the measurement procedures developed here pave the way for the future calibration program of the X-IFU thermal filters.
No takes yet. Share an insight, caveat, or question.
Sciortino et al. (2026) studied this question.