Answers are grounded in this paper and cite their sources.
A stability-indicating RP-HPLC method was developed and validated for the quantification of lamivudine in the bulk drug as well as tablet dosage forms using an Analytical Quality by Design (AQbD) approach. The Analytical Target Profile was mainly being defined, and critical analytical attributes were identified through the risk assessment and Failure Mode as well as Effects Analysis. It was found that the parameters of the critical method that were optimized after a 32 full factors design were the percentage of methanol in ammonium acetate buffer (pH 3.5) and the flow rate. Chromatography was used to extract the C18 column using methanol: ammonium acetate buffer (10:90, v/v; pH 3.5) at a rate of 0.9 mL/min and UV detection at 271 nm. Drug eluted at retention time of 3.9 min with the acceptable system suitability (theoretical plates 10533 ± 1600; tailing factor 0.98 ± 0.06). The method exhibited a high range of linearity (0-5 µg/mL, r² = 0.999), a high accuracy (99.21-100.84%), strong accuracy (% RSD = 2), and robustness. LOD and were 0.07 µg/mL and 0.18 µg/mL, respectively. The proposed AQbD-based on the RP-HPLC method is simple, robust, as well as suitable for the routine quality control as well as the stability assessment of lamivudine pharmaceutical formulations.
No takes yet. Share an insight, caveat, or question.
Jagdish V. Manwar*, Ganesh R. Pawar, Bhushan R. Gudalwar, Abhijit R. Raut, Priyanka P. Patil, Priyanka S. Patil (2026) studied this question.