Answers are grounded in this paper and cite their sources.
A molecularly imprinted electrochemical sensor (MIECS) tailored for trace-level quantification of losartan (LOS) was designed using a dual-functional monomer approach. A MIL-53(Fe)-N-doped carbon nanotube (MIL-53(Fe)-N-CNT) composite was first modified on the electrode to provide a large specific surface area and high electrical conductivity, thereby enhancing interfacial electron transfer and offering a high-surface-area platform for molecularly imprinted polymer (MIP) formation. The sensing MIP film was prepared through electropolymerization of p-aminobenzoic acid (p-ABA) and nicotinamide (NA) as co-monomers in the presence of LOS as the template. Compared with sensors prepared with single monomers, the combination of two monomers provided abundant functional groups and a synergistic recognition effect, significantly enhancing the specific recognition ability. Analytical performance was investigated via differential pulse voltammetry (DPV). The peak current varied linearly with the logarithm of LOS concentration across the range of 10-12 M to 10-6 M, yielding a detection limit of 2.29 × 10-13 M. The proposed sensor also exhibited good selectivity, repeatability, and stability. When applied to LOS tablet and plasma spiking tests, the recovery ranged from 96.3% to 107.0%. This work provides an alternative strategy for the selective and sensitive detection of losartan at trace levels.
No takes yet. Share an insight, caveat, or question.
Fu et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: