Answers are grounded in this paper and cite their sources.
Diffraction-limited synchrotron sources place stringent demands on wavefront-preserving X-ray optics. We report a fully in-house, closed-loop pipeline at the National Synchrotron Light Source II that links ion beam figuring, precision ex situ metrology, wavefront simulation, and in situ beam validation. The fabricated elliptical Kirkpatrick–Baez mirrors achieved 0.4 nm root-mean-square residual height error. At 12 keV, ptychography measured a 70.5 × 33.1 nm 2 (V × H) focal spot using an 8 × 7 µm 2 secondary source aperture, consistent with partially coherent Synchrotron Radiation Workshop simulations of 77 × 36 nm 2 . X-ray fluorescence imaging independently resolved approximately 30 nm features in a Siemens star. Back-propagation of the reconstructed probe further linked the measured wavefront to ex situ surface errors, establishing quantitative closure across fabrication, metrology, simulation, and beamline commissioning.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: