Electroencephalography (EEG) remains inaccessible in many health care settings, including low- and middle-income countries and point-of-care environments where trained EEG technologists are unavailable. Although low-cost EEG systems enable inexperienced users to acquire EEG, reliable acquisition also requires quality assurance of accurate electrode array positioning. We developed an artificial intelligence (AI)-enabled tool to address this need. We conducted a prospective randomized crossover validation study comparing AI-guided EEG cap fitting by inexperienced users with placement by an expert EEG technologist. A blinded expert quantified positioning error against International 10-20 System targets and independently rated overall placement as optimal, usable, or incorrect. Thirty participants completed 60 placement trials. Mean absolute positioning error was .938 cm with app-guided and .855 cm with expert placement. The mean paired difference was .084 cm (90% confidence interval = .024-.143), demonstrating noninferiority against the prespecified .5-cm margin. Clinically acceptable placement (optimal or usable) was achieved in 28 of 30 (93.3%) app-guided and 30 of 30 (100%) expert trials. AI-enabled quality assurance enabled inexperienced users to achieve electrode array positioning noninferior to expert placement, supporting its potential use in expanding access to point-of-care EEG.