To address the limited flexibility of neural models in handling tissue priors, instrument response functions (IRFs), and available channels, we propose the Prior- and IRF- Conditioned Transformer (PICT) for layer-resolved optical-property estimation using time-domain near-infrared spectroscopy. Training and evaluation used 6,000 Monte Carlo simulations, with local-phantom and public-data validation. Compared with IRF-ablated inference, PICT reduced normalized mean absolute error by 12.7–59.7% in one- to three-layer media. Compared with no-prior inference, PICT with known thicknesses reduced three-layer optical error by 19.5%. Compared with single-channel inference, four-channel inference reduced combination-averaged no-prior error by 19.8–47.7%. These results establish PICT as an effective, adaptable framework for layered optical-property estimation across tissue and measurement conditions.