Answers are grounded in this paper and cite their sources.
Building on partially coherent imaging theory, we develop a quantitative framework for full-field optical coherence tomography (FFOCT) that jointly accounts for spatial coherence and optical aberrations. We introduce performance metrics that separate lateral-resolution degradation from interferometric signal loss through a flux–shape factorization and an interferometric Strehl ratio. This factorization enables a rigorous analysis of the trade-off between resolution and signal level as a function of aberration amplitude and spatial coherence. Within this framework, we formulate an interferometric Maréchal criterion to evaluate aberration tolerance across coherence regimes. The model predicts a mode-dependent optimum in aberration tolerance. For defocus, astigmatism, and trefoil, the optimum occurs at an intermediate partial-coherence regime corresponding to a degree of spatial incoherence of approximately 0.8 on a scale from 0 (fully coherent) to 1 (limiting incoherent); coma instead remains best tolerated near the limiting incoherent regime. Finally, we provide practical FFOCT design guidelines for adaptive optics and computational correction strategies under realistic coherence conditions.
No takes yet. Share an insight, caveat, or question.
Loukili et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: