Ultraweak back-reflection measurement is essential in precision interferometric systems, but high-order echoes near the noise floor are vulnerable to noise fluctuations, sidelobes, and multipath interference. This study evaluates a processing workflow for high-order back-reflection echoes in balanced low-coherence interferometry. The complete workflow combines 800 Hz filtering for Echoes 1–4 and 100 Hz filtering for Echo 5 with measured-interval localization, multi-seed candidate selection, effective-amplitude correction, and covariance-based uncertainty evaluation. Experiments use a single 2.00 mm thick uncoated plane-parallel N-BK7 window at normal incidence. Echoes 2–4 provide the reflectivity calibration, and Echo 5 serves as the holdout order. Ten scans are processed under fixed optical alignment and nominal acquisition settings. The representative Echo 5 candidate has a narrowband-template amplitude signal-to-noise ratio (SNR) of 2.78 and a model-based reflectivity estimate of (1.66±0.37)×10 −10 . The expanded run-to-run relative uncertainty is 22.01% ( k =2), evaluated using the sample covariance of the final amplitude, width, and calibration parameters across the 10 scans. The complete workflow yields a mean Echo 5 reflectivity of 1.68×10 −10 and a run-to-run relative standard deviation (RSD) of 10.88%, compared with 49.91% for the defined baseline applied to the same 10 traces.