Electrochemical impedance spectroscopy (EIS) is susceptible to contributions from instrumentation, fixtures, grounding, shielding, and electromagnetic disturbances. This study examined enclosure and transformer-state sensitivity across two campaigns comprising 388 passive dummy-circuit spectra and 104 Samsung INR18650-25R cell spectra. Assessment combined relative complex-impedance deviations, replicate-based Type-A uncertainty, instrument accuracy-contour screening, and fixed-order, 24-RC KK-compatible residual analysis. With an energized but unloaded transformer, Campaign A dummy measurements yielded maximum deviations of 22.30% at 100 µV, 2.05% at 1 mV, and 0.274% at 10 mV, with the strongest response near 49.867 Hz. Campaign B reproduced the amplitude-dependent pattern across four repetitions per transformer state, but corresponding maxima were reduced to 5.31%, 0.460%, and 0.112%, demonstrating variable rather than fixed interference magnitude. Galvanostatic dummy deviations remained comparatively small. For the cell, plain-table transformer-state deviations were 0.163% at 49.867 Hz in Campaign A and 0.061–0.249% across Campaign B excitation settings; none of the Campaign B values exceeded the Type-A repeatability-based detection threshold. Campaign B maxima reached 2.01–2.19% at 79.5–100 kHz, where instrument and fixture contributions limit interpretation. Enclosure comparisons identified no consistently superior configuration under these conditions, while sequential cell-voltage evolution limited causal attribution. Thus, disturbance detectability depends on the complete measurement chain, and dummy-circuit findings cannot be quantitatively transferred to cylindrical-cell EIS.
No takes yet. Share an insight, caveat, or question.
Gulyás et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: