Robotic automation can raise laboratory throughput, but an undersized cell or weak coordination logic leads to capacity shortfalls and rework. We assess an automated water-sample preparation cell before commissioning with a deterministic model of route-dependent processing, position blocking until pickup, robot travel and completion-to-pickup waiting limits. An operational problem, which enforces material-handling feasibility, is distinguished from a hard-wait-constrained problem, which also limits every internal wait. A makespan certificate computed without an optimisation solver, constraint-programming schedules and bounds with time lags, four dispatching rules, their rollout and a rolling-horizon optimiser are compared on fifteen scenarios, structural alternatives, stochastic and three-day runs, and robot-workload variations. On the 230-sample design day, the best dispatching rules finish within 0.4% of the certificate, yet 44% of the samples exceed the 300 s limit, mostly before Method-2 extraction. A schedule satisfying every limit nevertheless completes in 23.2 h when admissions are timed against the extraction unit; a second extraction unit with a second GC-vial module nearly eliminates the violations under simple dispatching. Stochastic, hidden-emulsion and three-day runs preserve the dispatching-based diagnosis in the tested settings. If auxiliary robot work approaches the industrial movement accounting, the robot becomes the binding resource. The procedure can be re-instantiated for other laboratory cells.